/*! This file is auto-generated */ .wp-block-button__link{color:#fff;background-color:#32373c;border-radius:9999px;box-shadow:none;text-decoration:none;padding:calc(.667em + 2px) calc(1.333em + 2px);font-size:1.125em}.wp-block-file__button{background:#32373c;color:#fff;text-decoration:none} Q5EE Internal standard graph -- Data ... [FREE SOLUTION] | 91Ó°ÊÓ

91Ó°ÊÓ

Internal standard graph -- Data are shown below for chromatographic analysis of naphthalene (C10H8), using deuterated naphthalene (C10D8, in which D is the isotope 2H) as an internal standard. The two compounds emerge from the column at almost identical times and are measured by a mass spectrometer.

(a) Using a spreadsheet such asFigure 4-15, prepare a graph of Equation 5-12 showing peak area ratio(C10H8/C10D8)versus concentration ratio role="math" localid="1663559632352" ([C10H8]/C10D8) . Find the least-squares slope and intercept and their standard uncertainties. What is the theoretical value of the intercept? Is the observed value of the intercept within experimental uncertainty of the theoretical value?

(b) Find the quotientrole="math" localid="1663559638520" [C10H8]/[C10D8]for an unknown whose peak area ratio (C10H8/C10D8) is 0.652. Find the standard uncertainty for the peak area ratio.

(c) Here is why we try not to use 3-point calibration curves. For n = 3 data points, there is n - 2 = 1 degree of freedom, because 2 degrees of freedom are lost in computing the slope and intercept. Find the value of Student's for confidence and 1 degree of freedom. From the standard uncertainty in (b), compute the 95 % confidence interval for the quotient[C10H8]/[C10D8] . What is the percent relative uncertainty in the quotient[C10H8]/[C10D8]? Why do we avoid 3-point calibration curves?

Short Answer

Expert verified

a) The least squares slope and intercept and their standard deviations and the theoretical value was calculated.

b) The standard uncertainty uxfor the peak ratio is ux=0.0355

c) The uncertainty from a 3-point calibration is large that is 75% because students t is 12.706 where there is only 1 degree of freedom.

If we had just one more data point, which gives 2 degree of freedom,, 95 % confidence interval would decrease by a factor of 3.

Step by step solution

01

Concept used

LINEST:

The LINEST is a Microsoft excel function which is uses the least squares method to measure the statistics for a straight line and array describing that line it is built-in function that can be categorized.

LINEARITY:

It is measure of how well a calibration curve follows a straight line and displaying that the response is proportional to the quantity of the analyte.

R2=∑xi-xyi-y2∑xi-x2yi-y2

Where R means the square of the correlation coefficient.

Standard additions to one solution:

The graph can be drawn using the equation

plot Is+xVV0ÁåŸFunctiontoplotony-axisversus IX+IxXiSiVSV0ÁåŸfunctiontoplotonX-axis

Confidence Intervals:

The confidence interval is given by the equation:

Confidence interval x±tsn

=x±tux( since standard uncertaintylocalid="1663560297056" ux=s/n)

Where,

x is mean

s is standard deviation

t is Student's

n is number of measurements

uxis standard uncertainty

02

Calculate the least squares slope and intercept, and their standard deviations and the theoretical value

a)

A spreadsheet with x-axis and y-axis are

Every solution is made to constant volume.

Then [X] / [S] vs Ax/As is plotted.

The intercept of the graph is 0.0084 and um=0.0517, ub=0.0335; the theoretical value of the intercept is 0.

The observed value is less than one standard uncertainty away from 0, which lies with the experimental error of 0.

03

Calculate the standard uncertainty for the peak ratio

b)

A spreadsheet with x-axis and y-axis is

The x -intercept is calculated in B19 and B20 and its uncertainty C10H8/C10D8is 0.598.

The standard uncertainty isux=0.0355

04

Calculate the confidence interval and degree freedom

c)

A spreadsheet with x-axis and y-axis is

The -intercept is calculated in B19 and B 20 and its uncertainty C10H8/C10D8is 0.598.

The standard uncertainty is ux=0.0355

The students t for 95% confidence and 1 Degree freedom is 12.706. The 95% confidence for C10H8/C10D8is

=x±tux=0.598±12.7060.035=0.598±0.451

The relative uncertainty =0.4510.598=75%

The uncertainty from a 3-point calibration is large that is 75% because students t is 12.706, where there is only 1 degree of freedom.

If we had just one more data point which gives 2 degree of freedom, the 95% confidence interval would decrease by a factor of 3.

Unlock Step-by-Step Solutions & Ace Your Exams!

  • Full Textbook Solutions

    Get detailed explanations and key concepts

  • Unlimited Al creation

    Al flashcards, explanations, exams and more...

  • Ads-free access

    To over 500 millions flashcards

  • Money-back guarantee

    We refund you if you fail your exam.

Over 30 million students worldwide already upgrade their learning with 91Ó°ÊÓ!

One App. One Place for Learning.

All the tools & learning materials you need for study success - in one app.

Get started for free

Most popular questions from this chapter

Detection limit. In spectrophotometry, we measure the concentration of an analyte by its absorbance of light. A low-concentration sample was prepared and nine replicate measurements gave absorbances of 0.0047,0.0054,0.0062,0.0060,0.0046,0.0056,0.0052,0.0044, and 0.0058. Nine reagent blanks gave values of 0.0006,0.0012, 0.0022,0.0005,0.0016,0.0008,0.0017,0.0010, and 0.0011.

a) Find the absorbance detection limit with equation 5-3.

b) The calibration curve is a graph of absorbance versus concentration. Absorbance is a dimensionless quantity. The slope of the calibration curve is m=2.24x104M-1Find the concentration detection limit with Equation 5-5.

(c) Find the lower limit of quantitation with Equation 5-6.

Internal standard calibration curve. Figure 5-10 is a graph of Ax/Asversus[X]/[S]=(mol%vinylacetateunits)/(mol%ethyleneunits)=q/p in Reaction 5-13.

(d) From the uncertaintyubof the intercept, find the95%confidence interval for the intercept. Does this interval include the theoretical value of zero?

Detection limit. A sensitive chromatographic method was developed to measure sub-part-per-billion levels of the disinfectant by-products iodate(IO3-), chlorite(CIO2-), and bromate(BrO3-)in drinking water. As the oxyhalides emerge from the column, they absorption at267nm. For example, each mole of BrO3-+8Br-+6H+R3Br3-+3H2Obromate makes by the reaction

Bromate near its detection limit gave the following chromatographic peak heights and standard deviations (s). For each concentration, estimate the detection limit. Find the mean detection limit. The blank is 0 because chromatographic peak height is measured from the baseline adjacent to the peak. Because blank =0, relative standard deviation applies to both peak height and concentration, which are proportional to each other. Detection limit is 3 s for peak height or concentration.

Standard addition graph. Students performed an experiment like that in Figure 5-7 in which each flask contained 25.00mLof serum, varying additions of 2.640MNaCIstandard, and a total volume of 50.00mL.


(a) Prepare a standard addition graph and find [Na+]in the serum.

(b) Find the standard deviation and 95%confidence interval for [Na+].

What is the difference between an instrument detection limit and a method detection limit? What is the difference between robustness and intermediate precision?

See all solutions

Recommended explanations on Chemistry Textbooks

View all explanations

What do you think about this solution?

We value your feedback to improve our textbook solutions.

Study anywhere. Anytime. Across all devices.